Data sheet 90.2550 (90.2522) Page 1/4
2011-01-31/00311506
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow - Essex CM20 2DY, UK
Phone: +44 1279 63 55 33
Fax: +44 1279 63 52 62
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone: 315-697-JUMO
1-800-554-JUMO
Fax: 315-697-5867
Internet: www.jumo.us
RTD surface temperature probes
Temperature range from -50 to +260°C
With probes made of various materials
For round and level surfaces
Easy and quick installation
Low thermal mass
RTD surface temperature probes are used for measuring temperatures on closed pipework and/
or other round or level surfaces. Versions 902522/10 and 902522/11 can be attached to any sur-
face, for example a heating platen, via their mounting holes. The remaining versions avoid the
necessity for mechanical modification of the measurement location due to their easy installation
via cable ties or hose clamps.
An advantage to the indirect method of temperature measurement is that disruption of the flow
stream is avoided. Additionally, the life of the RTD is not shortened by the effects of either pres-
sure or chemicals.
The low thermal mass of the sensor means that there is very little influence of the sensor on the
measured object. Thermal conductance paste is available to improve heat transfer. Large dif-
ferences in temperature between the measured medium and the surrounding ambient air will
influence the measurement; in such cases, we recommend additional insulation around the sen-
sor.
The measurement insert is normally fitted with a Pt 100 temperature sensor as per DIN EN
60751, Class B in 2-wire circuit; versions with Pt 500 or Pt 1000 can also be supplied.
Technical Data
Connection Various cable termination types are available: tin-plated, with ferrules, contacts or multi-pole
connectors
Cable Silicone, ambient temperature -50 to +180°C
PTFE, ambient temperature -190 to +260°C
Kapton, ambient temperature -50 to +260 °C
Thermowell Stainless steel 1.4571 (equivalent to SS316 Ti), aluminum, plastic
Measuring insert Pt 100 temperature sensor, DIN EN 60 751, Class B, two-wire circuit
Vue de la page 0
1 2 3 4

Résumé du contenu

Page 1 - Technical Data

Data sheet 90.2550 (90.2522) Page 1/42011-01-31/00311506JUMO GmbH & Co. KGDelivery address: Mackenrodtstraße 14,36039 Fulda, GermanyPostal address

Page 2 - Dimensions

2011-01-31/00311506Data sheet 90.2550 (90.2522) Page 2/4JUMO GmbH & Co. KGDelivery address: Mackenrodtstraße 14,36039 Fulda, GermanyPostal address

Page 3 - Accessories

2011-01-31/00311506Data sheet 90.2550 (90.2522) Page 3/4JUMO GmbH & Co. KGDelivery address: Mackenrodtstraße 14,36039 Fulda, GermanyPostal address

Page 4 - Stock versions

2011-01-31/00311506Data sheet 90.2550 (90.2522) Page 4/4JUMO GmbH & Co. KGDelivery address: Mackenrodtstraße 14,36039 Fulda, GermanyPostal address